Pagina 1 - CENTRAL MONITOR

CNS-9701ACNS-9701JCNS-9701KCENTRAL MONITORCNS-97010634-001842C

Pagina 2 - Reader Comment Card

ii Service Manual CNS-9701(3) Technical information such as parts list, descriptions, calibration instructions or other information is available forq

Pagina 3 - Contents

4. DISASSEMBLY AND ASSEMBLY4.20 Service Manual CNS-97011. Remove the 4 screws which secure the rubber feet to the bottom of the recorderunit.Disasse

Pagina 4

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.214. Pull the recorder sub assy toward you until you can see 2 screws at the bottomof the recor

Pagina 5

4. DISASSEMBLY AND ASSEMBLY4.22 Service Manual CNS-97011. Remove the AC inlet socket from the recorder unit according to the “Removingthe AC Inlet S

Pagina 6

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.231. Remove the enclosure from the recorder unit according to the “Removing theRecorder Sub Ass

Pagina 7

4. DISASSEMBLY AND ASSEMBLY4.24 Service Manual CNS-9701Replacing the PlatenRoller/Bushings/Gear1. Open the paper magazine with the paper magazine re

Pagina 8 - C.6 Service Manual CNS-9701

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.251. Open the paper magazine with the paper magazine release lever. The thermalarray head and p

Pagina 9

4. DISASSEMBLY AND ASSEMBLY4.26 Service Manual CNS-9701This page is intentionally left blank.

Pagina 10 - WARRANTY POLICY

Service Manual CNS-9701 5C.1Section 5 MaintenanceTo Be Replaced Periodically ...

Pagina 11 - EMC RELATED CAUTION

5C.2 Service Manual CNS-9701This page is intentionally left blank.

Pagina 12

5. MAINTENANCEService Manual CNS-9701 5.1We recommend the periodic replacement of the following components accordingto the expected life span.To Be

Pagina 13

Service Manual CNS-9701 iiiEMC RELATED CAUTIONThis equipment and/or system complies with the International Standard IEC 60601-1-2 for electromagnetic

Pagina 14 - Power off

5. MAINTENANCE5.2 Service Manual CNS-9701Item Check Countermeasure against negative resultScreen position Check that the upper, lower, right and lef

Pagina 15 - Section 1 General

5. MAINTENANCEService Manual CNS-9701 5.3OperationItem Check Countermeasure against negative resultMouse Check that the mouse operation is normal. I

Pagina 16

5. MAINTENANCE5.4 Service Manual CNS-9701Laser PrinterItem Check Countermeasure against negative resultPrint key function Check that the trendgraph

Pagina 17 - Introduction

5. MAINTENANCEService Manual CNS-9701 5.5UPSItem Check Countermeasure against negative resultConnection to theUPSCheck that the power cords of the m

Pagina 18 - CAUTIONS

5. MAINTENANCE5.6 Service Manual CNS-9701Maintenance Check Sheet(Refer to the Maintenance section of the service manual for details.)Customer: _____

Pagina 19

5. MAINTENANCEService Manual CNS-9701 5.7UPSThe power cords of the main unit, display unit and recorder unit are connected to the UPSThe laser print

Pagina 20 - Service Parts

5. MAINTENANCE5.8 Service Manual CNS-9701This page is intentionally left blank.

Pagina 21 - Important Safety Information

Service Manual CNS-9701 6C.1Section 6 AdjustmentAdjusting the VL-971R Display Unit ...

Pagina 22

6C.2 Service Manual CNS-9701This page is intentionally left blank.

Pagina 23

6. ADJUSTMENTService Manual CNS-9701 6.1When a CRT display unit is used with a personal computer, adjustment of the CRTdisplay unit is not required.

Pagina 24

iv Service Manual CNS-9701In IEC 60601-1-2 Medical Electronic Equipment, Part 1: General Requirements for Safety, 2. CollateralStandard: Electromagne

Pagina 25

6. ADJUSTMENT6.2 Service Manual CNS-9701After performing the automatic adjustment, perform the following procedure.1. Check that the CNS system progr

Pagina 26 - NIBP Measurement

Service Manual CNS-9701 7C.1Section 7 Replaceable Parts ListMU-971RA/RJ/RK Main Unit (Vorax5000) ...

Pagina 27 - Maintenance

7C.2 Service Manual CNS-9701This page is intentionally left blank.

Pagina 28 - Specification

7. REPLACEABLE PARTS LISTService Manual CNS-9701 7.1When ordering parts or accessories from your nearest Nihon Kohden Corporationdistributor, please

Pagina 29

7. REPLACEABLE PARTS LIST7.2 Service Manual CNS-9701MU-971RA/RJ/RK Main Unit (Vorax5000)Index NK Parts Code Description1 635689 Power supply unit PC

Pagina 30 - Power Requirements

7. REPLACEABLE PARTS LISTService Manual CNS-9701 7.3aa165111079841232

Pagina 31 - Environment

7. REPLACEABLE PARTS LIST7.4 Service Manual CNS-9701MU-971RA/RJ/RK Main Unit (Vorax5001)Index NK Parts Code Description1 658691 Power supply unit PC

Pagina 32 - Panel Descriptions

7. REPLACEABLE PARTS LISTService Manual CNS-9701 7.5aa165111079841232

Pagina 33

7. REPLACEABLE PARTS LIST7.6 Service Manual CNS-9701VL-971R/RK Optional Display UnitIndex NK Code No. Qty Description1 646124 1 LCD assembly (backli

Pagina 34 - LCD Unit VL-971R/RK

7. REPLACEABLE PARTS LISTService Manual CNS-9701 7.712 V DC output20) Key switch assembly on the rear cover3) LCD control board10) Digital RGB

Pagina 35 - Recorder Unit WS-971R

Service Manual CNS-9701 vSymbol Description Symbol DescriptionPower ON Hard disk lampPower OFF ResetAttention, consult operator’smanualMouse socketSy

Pagina 36 - Composition

7. REPLACEABLE PARTS LIST7.8 Service Manual CNS-9701WS-971R Optional Recorder UnitIndex NK Code No. Qty Description1 648808 1 Power supply unit2 648

Pagina 37 - Connection Diagram

7. REPLACEABLE PARTS LISTService Manual CNS-9701 7.9Mains voltagePC socketRS232C interfaceboardLEDsRecorder sub assyRecording dataControl signalsSup

Pagina 38 - Central monitor main unit

7. REPLACEABLE PARTS LIST7.10 Service Manual CNS-9701This page is intentionally left blank.

Pagina 39 - Section 2 Troubleshooting

Service Manual CNS-9701 8C.1Section 8 Network SettingsNetwork Setting Flowchart ...

Pagina 40

8C.2 Service Manual CNS-9701This page is intentionally left blank.

Pagina 41

8. NETWORK SETTINGSService Manual CNS-9701 8.1This section describes how to set the following settings.• IP address• Install the laser printer (netw

Pagina 42

8. NETWORK SETTINGS8.2 Service Manual CNS-9701Network Setting Flowchart1. Set the IP address on each instrument.2. Connect the network system.3. Sta

Pagina 43 - Service Manual CNS-9701 2.3

8. NETWORK SETTINGSService Manual CNS-9701 8.3Each instrument in the network has an IP address for data communication.Duplicate IP addresses must no

Pagina 44 - 2.4 Service Manual CNS-9701

8. NETWORK SETTINGS8.4 Service Manual CNS-97013. Click “Local Area Connection”. The “Local Area Connection Status” dialogbox opens and shows the ne

Pagina 45 - Measurement Value

8. NETWORK SETTINGSService Manual CNS-9701 8.56. Enter the IP address and subnet mask.7. Click the “OK” button to finish setting.8. Restart Windows

Pagina 46 - Admitting/Discharging

vi Service Manual CNS-9701On VL-971R/RK LCD unitSymbol Description Symbol DescriptionPower onAttention, consult operator’smanualPower offOn WS-971R R

Pagina 47 - Service Manual CNS-9701 2.7

8. NETWORK SETTINGS8.6 Service Manual CNS-9701Install the laser printer (network printer) in the central monitor. Shut down thecentral monitor syst

Pagina 48 - Hemodynamics List Window

8. NETWORK SETTINGSService Manual CNS-9701 8.74. Click “Next” to start the “Add Printer Wizard”.5. Select “Use the following port”. Select the prin

Pagina 49 - ST Recall Window

8. NETWORK SETTINGS8.8 Service Manual CNS-97016. From Manufacturers, select HP. From Printers, select HP LaserJet 4000 SeriesPCL. (HP Laser Jet 41

Pagina 50 - ECG 12 Lead Analysis Window

8. NETWORK SETTINGSService Manual CNS-9701 8.99. Select “Do not share this printer” a10. Select “Yes” and click Next>.11. Click Finish.

Pagina 51 - Recording

8. NETWORK SETTINGS8.10 Service Manual CNS-970112. Files are copied and a test page is printed. After the test page is printed, clickOK.13. On the

Pagina 52 - Parameter Setup

8. NETWORK SETTINGSService Manual CNS-9701 8.1114. Shut down the window and restart it. Select restart and click OK.15. After the CNS-9701 system st

Pagina 53 - Power Problem

8. NETWORK SETTINGS8.12 Service Manual CNS-9701Connecting the Network SystemRefer to the Network and System installation guide and Section 3 “Connec

Pagina 54 - Display Problem

8. NETWORK SETTINGSService Manual CNS-9701 8.13You can see the network connection on the Maintenance window of the SystemSetup on the central monito

Pagina 55

8. NETWORK SETTINGS8.14 Service Manual CNS-9701To identify the monitors, each monitor has an instrument name (bed name/CNSname) and a group name. E

Pagina 56 - Sound Problem

8. NETWORK SETTINGSService Manual CNS-9701 8.15Assign an instrument name to the central monitor, beds on the multiple patientreceiver and bedside mo

Pagina 57 - Hard Disk Problem

Service Manual CNS-9701 1C.1Section 1 GeneralIntroduction ...

Pagina 58 - 2.18 Service Manual CNS-9701

8. NETWORK SETTINGS8.16 Service Manual CNS-9701Assigning a Group NameChecking the Bed Nameand Group Name SettingGroup nameAll monitors connected to

Pagina 59 - Error Messages

8. NETWORK SETTINGSService Manual CNS-9701 8.17One bedside monitor can be monitored on several central monitors, but the datacan only be stored on o

Pagina 60 - 2.20 Service Manual CNS-9701

8. NETWORK SETTINGS8.18 Service Manual CNS-9701NOTEFor a 8000 series bed whose data is stored in a CNS-9301/9302 centralmonitor, some waveforms cann

Pagina 61 - Service Manual CNS-9701 2.21

8. NETWORK SETTINGSService Manual CNS-9701 8.19Changing Settings of System Setup, Patient Setting and SetupBefore using the central monitor, confirm

Pagina 62 - 2.22 Service Manual CNS-9701

8. NETWORK SETTINGS8.20 Service Manual CNS-9701Adding an Instrument to the Network During System OperationAn instrument can be connected to a networ

Pagina 63 - VENT Alarm

Service Manual CNS-9701 9C.1Section 9 Connector PinAssignmentSockets at the Rear of the Main Unit ...

Pagina 64 - 2.24 Service Manual CNS-9701

9C.2 Service Manual CNS-9701This page is intentionally left blank.

Pagina 65 - Messages From Central

9. CONNECTOR PIN ASSIGNMENTService Manual CNS-9701 9.1Sockets at the Rear of the Main UnitCAUTIONConnect only the specified instruments to the connec

Pagina 66 - Hard Disk Replacement Flow

9. CONNECTOR PIN ASSIGNMENT9.2 Service Manual CNS-9701PROTECT KEY SocketRECORDER SocketPin No. Signal1 Data Carrier Detect2 Transmit Data3 Receive Da

Pagina 67

1C.2 Service Manual CNS-9701Environment ...

Pagina 68 - 2.28 Service Manual CNS-9701

Service Manual CNS-9701 1.11. GENERALIntroductionThis service manual provides useful information to qualified service personnel tounderstand, troubl

Pagina 69 - Service Manual CNS-9701 2.29

1.2 Service Manual CNS-97011. GENERALGeneral Information on ServicingNote the following information when servicing the monitor.WARNINGTo avoid the p

Pagina 70 - PASSWORD

Service Manual CNS-9701 1.31. GENERALEnvironmental SafeguardsDepending on the local laws in your community, it may be illegal todispose of the lithi

Pagina 71 - Service Manual CNS-9701 2.31

Model: CNS-9701Manual code no.: 0634-001842CReader Comment CardWe welcome your comments about this manual. Your comments and suggestions help us impr

Pagina 72 - Shutdown Temperature Disabled

1.4 Service Manual CNS-97011. GENERALMaintenance Equipmentsand ToolsService Policy, Service Parts and Patient Safety ChecksService Policy Our techni

Pagina 73

Service Manual CNS-9701 1.51. GENERALImportant Safety InformationWARNING••••• Never use the central monitor in the presence of any flammableanesthet

Pagina 74 - 3C.2 Service Manual CNS-9701

1.6 Service Manual CNS-97011. GENERALInstallationWARNINGInstall the central monitor and external instruments outside thepatient environment (IEC 606

Pagina 75 - Motherboard (by A-OPEN)

Service Manual CNS-9701 1.71. GENERALNetworkWARNING••••• Connect the central monitor to the network as specified in theNetwork and System Installati

Pagina 76 - CD-ROM Drive (by SAMSUNG)

1.8 Service Manual CNS-97011. GENERALTurning the Power On/OffWARNING••••• Do not turn off the display to which the sound cable is connected.Otherwis

Pagina 77 - Fan for Chassis

Service Manual CNS-9701 1.91. GENERALTemporary DischargeWARNINGWhen a temporarily discharged patient comes back, resumemonitoring for the patient.

Pagina 78 - 3.4 Service Manual CNS-9701

1.10 Service Manual CNS-97011. GENERALECG MonitoringCAUTION••••• If there is any doubt about the arrhythmia analysis, make the centralmonitor relear

Pagina 79

Service Manual CNS-9701 1.111. GENERALMaintenanceCAUTION••••• Before maintenance (inspection, cleaning, disinfection), turn thecentral monitor power

Pagina 80 - 4C.2 Service Manual CNS-9701

1.12 Service Manual CNS-97011. GENERALNumber of patients on the display 16 patients maxWith overview bed function, 16 patients plusone other patient

Pagina 81 - Before You Begin

Service Manual CNS-9701 1.131. GENERALArrhythmia: ASYSTOLE, V FIB, V TACHY, EXT TACHY,EXT BRADY, VPC RUN, COUPLET, MULTIFORM, EARLY VPC, BIGEMINY, T

Pagina 82 - Disassembling the Main Unit

CONTENTSService Manual CNS-9701 C.1ContentsGENERAL HANDLING PRECAUTIONS ... iWA

Pagina 83

1.14 Service Manual CNS-97011. GENERALFull disclosure data (only when the QP-973P Full Disclosure Filing Program Kit isinstalled.)ECG 12Lead analysi

Pagina 84

Service Manual CNS-9701 1.151. GENERALOperating temperature 10 to 35°COperating humidity 30% to 80% (no condensing)Operating atmospheric pressure 70

Pagina 85

1.16 Service Manual CNS-97011. GENERALCentral Monitor Main UnitMU-971RA/RJ/RKFront PanelWARNINGWhile the central monitor is on, do not insert any CD

Pagina 86

Service Manual CNS-9701 1.171. GENERALNOTE••••• Do not remove the covers from the unused sockets on the rear panelof the central monitor main unit.•

Pagina 87

1.18 Service Manual CNS-97011. GENERALLCD Unit VL-971R/RKFrontRearWARNINGDo not turn off the display to which the sound cable is connected.Otherwise

Pagina 88

Service Manual CNS-9701 1.191. GENERALRecorder Unit WS-971RPOWER lampLights when the power is on.CHECK RECORDER lampLights when communicationbetween

Pagina 89

1.20 Service Manual CNS-97011. GENERALCompositionOptionsMU-971RA Central monitor main unit for A versionMU-971RJ Central monitor main unit for J ver

Pagina 90

Service Manual CNS-9701 1.211. GENERALConnection DiagramMouseKeyboardLCD unit (VL-971R/RK)Central monitor main unit(MU-971RA/RJ/RK)Recorder unitWS-9

Pagina 91 - Replacing the Graphic

1.22 Service Manual CNS-97011. GENERALMain displaySub displayLCD unit (VL-971R/RK)Central monitor main unit(MU-971RA/RJ/RK)When using two displays

Pagina 92

Service Manual CNS-9701 2C.1Section 2 TroubleshootingGeneral ...

Pagina 93 - Replacing the LCD Assembly

CONTENTSC.2 Service Manual CNS-9701Safety ...

Pagina 94

2C.2 Service Manual CNS-9701Other ... 2.25H

Pagina 95 - Removing the Display Unit

2. TROUBLESHOOTINGService Manual CNS-9701 2.1Use the troubleshooting tables to locate, identify, and solve a problem in theinstrument. The problems

Pagina 96 - Touch screen

2. TROUBLESHOOTING2.2 Service Manual CNS-9701How to use the troubleshooting table1. Determine which troubleshooting table to use.2. In the “Problem”

Pagina 97

2. TROUBLESHOOTINGService Manual CNS-9701 2.3SystemTrouble Possible Cause/Criteria ActionThe touch screen keys do notfunction.The pressed position a

Pagina 98 - 9) Speaker

2. TROUBLESHOOTING2.4 Service Manual CNS-9701NetworkTrouble Possible Cause/Criteria ActionData is deleted when the centralmonitor power is turned of

Pagina 99 - Replacing the Touch

2. TROUBLESHOOTINGService Manual CNS-9701 2.5Measurement ValueAlarmTrouble Possible Cause/Criteria ActionWaveform too large or too smallQRS detectio

Pagina 100 - Sub Assy

2. TROUBLESHOOTING2.6 Service Manual CNS-9701Admitting/DischargingTrouble Possible Cause/Criteria ActionCannot transfer a patient’s data.Cannot chan

Pagina 101 - Removing the AC Inlet

2. TROUBLESHOOTINGService Manual CNS-9701 2.7All Beds Screen/Individual Bed ScreenTrouble Possible Cause/Criteria ActionNumericdataandmessagesdisapp

Pagina 102

2. TROUBLESHOOTING2.8 Service Manual CNS-9701Trend WindowTabular Trend WindowHemodynamics List WindowTrouble Possible Cause/Criteria ActionTrend dat

Pagina 103 - Replacing the LED

2. TROUBLESHOOTINGService Manual CNS-9701 2.9Arrhythmia Recall WindowST Recall WindowTrouble Possible Cause/Criteria ActionArrhythmia recall files a

Pagina 104 - Roller/Bushings/Gear

CONTENTSService Manual CNS-9701 C.3Laser Printer ... 2

Pagina 105

2. TROUBLESHOOTING2.10 Service Manual CNS-9701Full Disclosure WindowECG 12 Lead Analysis WindowTrouble Possible Cause/Criteria ActionThe central mon

Pagina 106 - 4. DISASSEMBLY AND ASSEMBLY

2. TROUBLESHOOTINGService Manual CNS-9701 2.11Report WindowRecordingTrouble Possible Cause/Criteria ActionReport printing startsautomatically.The re

Pagina 107 - Section 5 Maintenance

2. TROUBLESHOOTING2.12 Service Manual CNS-9701Overview Bed ScreenParameter SetupTrouble Possible Cause/Criteria ActionSome waveforms are notdisplaye

Pagina 108 - 5C.2 Service Manual CNS-9701

2. TROUBLESHOOTINGService Manual CNS-9701 2.13Power ProblemProblem Possible Cause ActionAC power is not obtained from thepower cord.Check that the A

Pagina 109 - To Be Replaced Periodically

2. TROUBLESHOOTING2.14 Service Manual CNS-9701Display ProblemCAUTIONTo replace the hard disk with a new one, refer to “Hard Disk Flow” inthis sectio

Pagina 110 - External and Connection

2. TROUBLESHOOTINGService Manual CNS-9701 2.15*1 The VL-971R display unit is adjusted at the factory so that the display unit accepts the RGB signal

Pagina 111 - Data Storage

2. TROUBLESHOOTING2.16 Service Manual CNS-9701Touch Screen ProblemSound ProblemProblem Possible Cause ActionThe RS232C straight cable haspoor contac

Pagina 112 - Recorder

2. TROUBLESHOOTINGService Manual CNS-9701 2.17Keyboard ProblemMouse ProblemHard Disk ProblemProblem Possible Cause ActionThe keyboard cable is loose

Pagina 113 - Service Manual CNS-9701 5.5

2. TROUBLESHOOTING2.18 Service Manual CNS-9701CD-ROM Drive or Floppy Disk Drive Problem2-channel Recorder Unit ProblemProblem Possible Cause ActionT

Pagina 114 - Maintenance Check Sheet

2. TROUBLESHOOTINGService Manual CNS-9701 2.19Error MessagesMessages From BedsideMonitors/Multiple PatientReceiverError messages are categorized int

Pagina 115

CONTENTSC.4 Service Manual CNS-9701Cleaning the Thermal Array Head/Paper Sensor ... 4.25Operation

Pagina 116 - 5.8 Service Manual CNS-9701

2. TROUBLESHOOTING2.20 Service Manual CNS-9701Parameter MessageModule FailureToo Many ParamsConnector OffProbe OffProbe DisconnectProbe BrokenNo Pul

Pagina 117 - Section 6 Adjustment

2. TROUBLESHOOTINGService Manual CNS-9701 2.21Parameter MessageModule FailureToo Many ParamsCheck LabelConnector OffCheck SensorOut of RangeZero Imb

Pagina 118 - 6C.2 Service Manual CNS-9701

2. TROUBLESHOOTING2.22 Service Manual CNS-9701Parameter MessageUnit FailureToo Many ParamsCheck LabelConnector OffCheck CatheterCatheter LocationMea

Pagina 119 - Automatic Adjustment

2. TROUBLESHOOTINGService Manual CNS-9701 2.23Parameter MessageModule FailureCom ErrorAlarmtcPO2/tcPCO2ConnectorOffParameter MessageModule FailureTo

Pagina 120 - Manual Adjustment

2. TROUBLESHOOTING2.24 Service Manual CNS-9701Trouble Possible Cause/Criteria ActionSIGNAL LOSS Weak electric field strength of awireless telemeter.

Pagina 121 - Service Manual CNS-9701 7C.1

2. TROUBLESHOOTINGService Manual CNS-9701 2.25Messages From CentralMonitorWS-971R Recorder UnitTrouble Possible Cause/Criteria ActionRECORDER DOOR O

Pagina 122 - 7C.2 Service Manual CNS-9701

2. TROUBLESHOOTING2.26 Service Manual CNS-9701Hard Disk Replacement FlowWhen you order a new hard disk for replacement, the new hard disk has the OS

Pagina 123

2. TROUBLESHOOTINGService Manual CNS-9701 2.2712. Install the network printer drive into the central monitor under Windows 2000.Refer to “Installing

Pagina 124 - 7.2 Service Manual CNS-9701

2. TROUBLESHOOTING2.28 Service Manual CNS-9701Virus Warning DisabledCPU Internal Cache EnabledExternal Cache EnabledCPU L2 Cache ECC checking Enable

Pagina 125 - Service Manual CNS-9701 7.3

2. TROUBLESHOOTINGService Manual CNS-9701 2.29On-Chip Primary PCI IDE EnabledOn-Chip Secondary PCI IDE EnabledIDE Primary Master PIO AutoIDE Primary

Pagina 126

CONTENTSService Manual CNS-9701 C.5Section 9 Connector Pin Assignment ... 9C.1Sockets at the Rear of the

Pagina 127 - Service Manual CNS-9701 7.5

2. TROUBLESHOOTING2.30 Service Manual CNS-9701Reset Configuration Data DisableResources Controlled by Auto (ESCD)PCI/VGA Palette Snoop DisabledAssig

Pagina 128 - 7.6 Service Manual CNS-9701

2. TROUBLESHOOTINGService Manual CNS-9701 2.31Dram Read Thermal Mgmt DisabledSystem BIOS Cacheable DisabledVideo BIOS Cacheable DisabledVideo RAM Ca

Pagina 129 - Service Manual CNS-9701 7.7

2. TROUBLESHOOTING2.32 Service Manual CNS-9701Reset Configuration Data DisabledResources Controlled By Auto (ESCD)PnP/PCI ConfigurationPCI/VGA Palet

Pagina 130 - 7.8 Service Manual CNS-9701

Service Manual CNS-9701 3C.1Section 3 Board/Unit DescriptionPower Supply Unit (by NIPRON) ...

Pagina 131 - Service Manual CNS-9701 7.9

3C.2 Service Manual CNS-9701This page is intentionally left blank.

Pagina 132 - 7.10 Service Manual CNS-9701

3. BOARD UNIT/DESCRIPTIONService Manual CNS-9701 3.1The power supply unit accepts AC line voltages of 100 to 240 V AC. The fan in thepower supply un

Pagina 133 - Section 8 Network Settings

3. BOARD/UNIT DESCRIPTION3.2 Service Manual CNS-9701256 MB Memory DIMM S-DRAM or DDR266One chip 256 MB memory for working area is used. In future, a

Pagina 134 - 8C.2 Service Manual CNS-9701

3. BOARD UNIT/DESCRIPTIONService Manual CNS-9701 3.3The front panel, right and left side panels and top cover are included in the chassis.Note that

Pagina 135

3. BOARD/UNIT DESCRIPTION3.4 Service Manual CNS-9701This page is intentionally left blank.

Pagina 136 - Network Setting Flowchart

Service Manual CNS-9701 4C.1Section 4 Disassembly and AssemblyBefore You Begin ...

Pagina 137 - Address of the Central

CONTENTSC.6 Service Manual CNS-9701This page is intentionally left blank.

Pagina 138

4C.2 Service Manual CNS-9701This page is intentionally left blank.

Pagina 139 - 8. Restart Windows 2000

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.1Before You BeginWarnings and CautionThe procedures in this section tell how to remove, replace

Pagina 140

4. DISASSEMBLY AND ASSEMBLY4.2 Service Manual CNS-9701Before performing the following procedure, shut down the main unit according to theoperator’s

Pagina 141

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.32. Remove the 3 screws which secure the 2 graphic boards and serial interface boardto the chas

Pagina 142

4. DISASSEMBLY AND ASSEMBLY4.4 Service Manual CNS-97015. Disconnect each power supply cable from the hard disk and floppy disk drive.6. Remove the d

Pagina 143 - 11. Click Finish

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.58. Disconnect the connection cable from the front panel which has the power switch,power indic

Pagina 144

4. DISASSEMBLY AND ASSEMBLY4.6 Service Manual CNS-970110. Remove the 6 screws which secure the motherboard to the chassis. Remove themotherboard fro

Pagina 145 - Service Manual CNS-9701 8.11

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.7NOTE in Assembling the Main UnitTo attach the CPU fan to the motherboard, latch the 2 short ho

Pagina 146 - Start the System

4. DISASSEMBLY AND ASSEMBLY4.8 Service Manual CNS-970115. Pull the lever of the CPU socket upward and remove the CPU from themotherboard.NOTE in Ass

Pagina 147

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.913. Release the 4 hooks of the CPU fan by pulling the 2 levers near the CPU fanupward. Remove

Pagina 148

Service Manual CNS-9701 iGENERAL HANDLING PRECAUTIONSThis device is intended for use only by qualified medical personnel.Use only Nihon Kohden approv

Pagina 149 - Registering Group Names

4. DISASSEMBLY AND ASSEMBLY4.10 Service Manual CNS-9701Replacing the LithiumBattery on theMotherboardAfter removing the motherboard from the main un

Pagina 150 - Checking the Bed Name

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.11Replacing the Hard DiskReplacing the FansBefore disassembling the main unit to replace the ha

Pagina 151 - Selecting the Monitored Beds

4. DISASSEMBLY AND ASSEMBLY4.12 Service Manual CNS-9701Replacing the Floppy DiskDriveRemove the screw which secures the drive to the disk case and r

Pagina 152

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.13To disassemble the display unit, the following tools are necessary.- Phillips screwdriver for

Pagina 153

4. DISASSEMBLY AND ASSEMBLY4.14 Service Manual CNS-97014. Remove the 12 nuts which secure the LCD assembly to the display unitchassis.5. Slightly pu

Pagina 154

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.15Removing the Display UnitChassis6. Remove the 4 screws (3 mm in dia.) and 6 screws (4 mm in d

Pagina 155 - Assignment

4. DISASSEMBLY AND ASSEMBLY4.16 Service Manual CNS-9701The following picture shows the cables which must be connected to the correspondingboards or

Pagina 156 - 9C.2 Service Manual CNS-9701

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.17CAUTION in Assembling the Display UnitCheck that the following three things are performed whe

Pagina 157 - TOUCH1/TOUCH2 Socket

4. DISASSEMBLY AND ASSEMBLY4.18 Service Manual CNS-9701- Ground wire from the baseA ground wire (green/yellow) which comes from the base must beconn

Pagina 158 - VIDEO1/VIDEO2 Socket

4. DISASSEMBLY AND ASSEMBLYService Manual CNS-9701 4.19When replacing the touch screen with a new one, the following parts should also bereplaced wi